Thermo TVA 2020

Thermo TVA 2020 Portable VOC Monitor (PID and FID)

Quickly detect organic and inorganic compounds in demanding applications such as Method 21 monitoring and site remediation with this lightweight, intrinsically safe portable detector. The Thermo Scientific™ TVA2020 Vapor Analyzer can be configured as flame ionization detector (FID) or dual detection technology (FID/PID-photo ionization detector) to offer fast response times. Key features include reduced response times, increased user-serviceability and Bluetooth™ connectivity.

Product Size -
Accuracy FID Instrument–±10% of reading or ±1.0 ppm, whichever is greater, from 1.0 to 10,000 ppm 
PID Instrument–±20 % of reading or ±0.5 ppm, whichever is greater, from 0.5 to 500 ppm
Components Measured Most organic and inorganic compounds
Computer Requirements USB port, Windows XP operating system or higher
Concentration Measurement Range FID: 0-30,000 ppm (methane)PID: 0-2,000 ppm (isobutylene)
Data Logging Averaging Periods Auto mode - 1 per second to 1 per 999 minutes, user selectable
VOC or FE Mode - 2 to 30 seconds, user selectable

Auto mode - 1 per second to 1 per 999 minutes, user selectable
VOC or FE Mode - 2 to 30 seconds, user selectable
Description TVA2020 Toxic Vapor Analyzer
Flow Rate 1L/min., nominal at sample probe inlet
Height (English) 11.5 in
Height (Metric) 29.2 cm
Hertz 50/60
Inputs BluetoothTM, USB
Depth (English) 4 in.
Depth (Metric) 10.2vm
Outputs BluetoothTM, USB
Repeatability FID Instrument - 2% at 500 ppm of methane 
PID Instrument - 1% at 100 ppm of isobutylene
Serial Interface USB
Voltage 100-230 V
Weight (English) FID only—9.2 pounds
Dual—9.4 pounds

FID only—9.2 pounds
Dual—9.4 pounds
Weight (Metric) FID only— 4.17 kg
Dual— 4.26 kg

FID only— 4.17 kg
Dual— 4.26 kg
Width (English) 9 in
Width (Metric) 22.9 cm
Battery Life The battery operating time is 10 hours minimum at 0 °C (32 °F). Fully charged in less than 10 hours.
Battery Type Lithium ion
Calibration Kit Part Number for FID: CR009UY 
Part Number for dual detector: CR012UH
Detector Type FID or PID/FID
Fuel 99.99% Hfor FID
Min. Detectable Limit The minimum detectable level is defined as seven times the standard deviation of peak-to-peak noise.
PID Instrument - 0.5 ppm of isobutylene
FID Instrument - 0.5 ppm of methane

The minimum detectable level is defined as seven times the standard deviation of peak-to-peak noise.
PID Instrument - 0.5 ppm of isobutylene
FID Instrument - 0.5 ppm of methane
Portable Operation Time Minimum 10 hours
Response Time PID Instrument - Less than 3.5 seconds for 90% of final value, using 500 ppm of isobutylene 
FID Instrument - Less than 3.5 seconds for 90% of final value, using 10,000 ppm of methane
Temperature (English) Range 14° F - 113° F
Temperature (Metric) Range -10°C - +45°C
Warranty One year from date of shipment