2BTech-202 Ozone

2BTech-202 Ozone Single Beam

The Model 202 Ozone Monitor is designed to enable accurate and precise (±1.5 ppb) measurements of ozone ranging from a few ppb to 100,000 ppb (0-100 ppm) based on the well established technique of UV absorbance at 254 nm. The Model 202 is light weight (2.1 kg.) and has a power consumption of only 4 watts. The Model 202 has been certified by the EPA as a Federal Equivalent Method (FEM): EQOA-0410-190  The Model 202 has been used for the past decade in ozone monitoring networks of the U.S. National Park Service and National Forest Service. It also is widely used by NOAA, EPA, Environment Canada and the British Antarctic Survey for studies of ground-level ozone.    

Measurement Principle UV Absorption at 254 nm
Federal Equivalent Method (FEM) Yes, EQOA-0410-190
Linear Dynamic Range 1.5 ppb to 250 ppm
Resolution 0.1 ppb
Precision (1σ; rms noise) Greater of 1.5 ppb or 2% of reading
Accuracy Greater of 1.5 ppb or 2% of reading
Limit of Detection (2σ) 3.0 ppb
NIST-Traceable Calibration Yes
Measurement Interval 10 s (Data averaging options: 10 s, 1 min, 5 min, 1 hr)
Flow Rate (nominal) ~1 Liter/min
Flow Rate Requirement >0.6 L/min
Baseline Drift < 2 ppb/day
< 5 ppb/year
Sensitivity Drift < 1%/day
< 3%/year
Measurement Time, Frequency 10 s, 0.1 Hz
Response Time, 100% of Step Change 20 s, 2 points
Averaging Times 10 s, 1 min, 5 min, 1 hr
Data Logger Capacity 14,336 lines (10 s avg. = 1.4 days; 1 min avg = 10 days; 5 min avg = 1.4 mo;
1 hr avg = 1.6 yr)
Ozone Units ppb, ppm, pphm, µg m-3, mg m-3
Pressure Units mbar, torr
Temperature Units °C, K
T and P Corrected Yes
Operating Temperature Range 0 to 50 °C; -20 to 50 °C with low temperature modifications (rotary vane pump and lamp heater)
Operating Altitude Range ~0-13.5 km (150-1,013 mbar) standard; ~0-25 km (30-1,013 mbar) with upgraded pressure sensor
Power Requirement;
Supplied by battery or 110/220 VAC Power Pack
12 V dc or 120/240 V ac, 335 mA, 4.0 watt (2.9 watt with cell heater unplugged)
Size 3.5 x 8.5 x 11 inches (9 x 21 x 29 cm)
Weight 4.7 lb (2.1 kg); 1.6 lb (0.7 kg) without instrument case
Analog Inputs for Internal Logging of Other Instruments 3 Analog Inputs, 0-2.5 V; For example could log external T, P and RH
Data Outputs RS232, 0-2.5 V Analog, LCD Display: (Optional 4-20 mA Current, External USB Converter; request quote)
Data Transfer Baud Rates 1200, 2400, 4800, 19200
Output Ranges User Defined Scaling Factor in Menu
DewLine™ Yes
Backup Air Pump Yes
Flow Meter Yes
Options GPS, Flash Card Memory, 4-20 mA Current Output; Lamp Heater; Rotary Vane Air Pump; Long-Life External Air Pump; Rack Mount Case
Upgradable to Model 205 Dual Beam Yes
I want to measure Ozone [O3]
Products Gas Monitors,Portable Monitors
Applications Outdoor Air Quality,Indoor Air Quality