Thermo - ADR 1500
Thermo ADR - Indicative PM Monitor
The ADR is a nephelometer designed to measure TSP, PM10 or PM2.5 by use of a size selective inlet. It features the same technoogy as the Thermo PDR housed in an IP65 cabinet with optional visual beacon. Ideal for construction sites and monitoring for dust around quarries and open cast mining.
Product Size | - |
---|---|
2-sigma Values | ±0.5% of reading or ±0.0015mg/m3, whichever is larger, for 10-second averaging time |
Accuracy | ±5% of reading ±precision) traceable to SAE Fine test dust |
Aerodynamic Particle Cut-Point Range | 1.0 to 10µm, with optional cyclone accessories |
Alarm Averaging Time | Real-time (1 - 60 seconds) or STEL (15 minutes) |
Alarm Output | Load impedance >100kw , Alarm ON = short to ground, Alarm OFF = open |
Concentration Measurement Range | 0.001 to 400mg/m³ (auto-ranging) |
Data Logging Averaging Periods | 1 second to 1hour |
Description | Area Dust Monitor |
Flow Rate Range | 1.0 to 3.5L/min. |
Height (English) | 21 in. |
Height (Metric) | 431mm |
Hertz | 50/60Hz |
Internal Battery Run Time | 1.2L/min. (heater and alarm constantly on) >24-hour run time |
Depth (English) | 17 in. |
Depth (Metric) | 215mm |
Logged Data | Averaged concentrations, temperature, RH, barometric pressure, time/date, and data point number |
Optional Cyclones | GK 2.05 (Red) primarily for PM4 through PM10, SCC 1.062 (blue) primarily for PM1 through PM4 |
Particle Size Range | 0.1 to 10µm |
Precision | ±2%, for 1-second averaging time |
Real-time Analog Signal | 0 to 2V and 4 to 20mA, with selectable full scale ranges between 0.1 and 400mg/m3; 0-0.1, 0-0.4, 0-1.0, 0-4.0, 0-10, 0-40, 0-100 and 0-400, mg/m3 |
Real-time Digital Signal | Concentration, flow, temperature, relative humidity, barometric pressure, time, date |
Repeatability | ±0.005mg/m3, for 1-second averaging time |
Resolution | 0.1µg/m? |
RH correction | ±0.2% or reading or ±0.0005mg/m3, whichever is larger, for 60-second averaging time |
Run Summary | Site number, average and max. concentrations, time/date of maximum, number of logged points, start time/date, elapsed time (run duration), averaging (logging) period, calibration factor and tag number |
Scattering Coefficient Range | 1.1 x 10-6 to 0.6m-1 (approximately) @ l = 880nm |
Serial Interface | High speed, USB/RS-232 (reserved for wireless applications), 19,200 baud |
Temperature (English) Operating | 14° to 122°F |
Temperature (Metric) Operating | 10° to 50°C |
Voltage | 100-240VAC, 12/24VDC for solar or auxiliary battery options |
Weight (English) | 28.5 lb. |
Weight (Metric) | 12.9kg |
Width (English) | 17 in. |
Width (Metric) | 431mm |
Item Description | Thermo Scientific ADR1500 |
I want to measure | Particulate Matter [PM10],Particulate Matter [PM2.5],Particulate Matter Total [TSP] |
---|---|
Products | Particle Monitors |
Applications | Construction Dust |